9 system calibration, Chapter 9 system calibration, Chapter 9, system calibration – Bio-Rad Image Lab™ Software User Manual

Page 187: 9system calibration

Advertising
background image

User Guide | 187

9

System Calibration

When your imager is installed, instrument calibration is performed with a calibration

wizard. See the installation guide in your Image Lab™ upgrade kit for detailed

instructions.

The instrument calibration wizard performs several procedures required to automate

the system and prevent focus problems. Each of these calibrations affects your

system as follows:

Focus Calibration — this calibration allows automated focus settings at

any zoom point, using a (patent pending) software algorithm. Therefore,

your focus remains correct whether you view an entire sample or an area of

interest.

Focus Calibration with Height Offset — this calibration takes the tallest

of the available conversion screens into account, and extrapolates values

for the others, so that focus remains optimal, for whichever screen is used.

Dark Reference Image — this calibration determines and corrects any

background signal present in your imager

UV Flat Field Calibration — this calibration generates the flat field correction

profiles needed for the UV light source. Because of this calibration, your

images have more accurate quantity reporting and backgrounds of even

intensity.

Lens Flat Field Calibration — this calibration corrects for the intensity roll-

off inherent in any lens.

White Conversion Screen Calibration — this calibration generates a flat

field correction profile needed for the white light conversion screen.