Detect the bands – Bio-Rad Image Lab™ Software User Manual

Page 148

Advertising
background image

6 | Analyzing Images

148 | ChemiDoc MP Imaging System with Image Lab Software

Detect the Bands

Use the band detection tool to detect the bands in the channels. If you are using the

total lane protein method to normalize your data, it is recommended that you do not

use the band detection tool on the normalization channel, because this will generate

large amounts of superfluous data in the analysis table. For more information, see

Detecting Bands on page 138

.

If you are using a housekeeping protein to calculate the normalization factor, you

must isolate this protein in your image. Remove all bands other than the

housekeeping protein from your image. The normalization channel should have only

one detected band in each lane. Use the Delete Bands tool to exclude all other

bands from the calculation.