Bio-Rad Image Lab™ Software User Manual

Page 77

Advertising
background image

Creating a Protocol

User Guide | 77

SAM is used to simplify the capture of a good image from a chemiluminescent

sample. This sample type often requires long integration times to obtain an

image that represents the best range of signal.

Rather than manually acquiring a series of independent images with different

imaging times, SAM presents a series of cumulative images with progressively

greater signal in each image. To run SAM you must estimate the shortest and

longest times expected to generate an image with the appropriate signal

intensity. You then decide how many total images to acquire in this window of

time.

For example, if the minimum expected time to image the sample is 1 minute

and the maximum is 5 minutes, these values are entered (in sec) in the setup

window. By entering 5 in the Total number of images field, three images will be

acquired between the first and last images.

To use SAM, click Signal Accumulation Mode and select Setup to display the

Signal Accumulation Setup dialog box.

In this example, the bar in the Signal Accumulation Setup dialog box indicates

that images will be acquired at 1 minute intervals, starting at 1 minute and

ending at 5 minutes. The second 1 minute image is added to the first 1 minute

image, and the final image is the result of integrating these two images. The

third 1 minute image is added to the previous image, and so on, until the last

image is presented.

Advertising
This manual is related to the following products: