Fuse replacement – Bio-Rad Image Lab™ Software User Manual

Page 215

Advertising
background image

Fuse Replacement

User Guide | 215

Fuse Replacement

Always unplug the instrument before changing or checking the fuses.

This unit is protected by two fuses (5 x 20 mm, 2 A Slo-Blo). The fuses are located in

fuse holders housed in the power entry module. This module is located on the right

side of the back of the universal hood.

To replace the fuses

1.

Unplug the main power cable from the power outlet.

2.

Use a flat screwdriver to turn the slotted front of each fuse holder
counterclockwise; the holder pops out so you can extract the fuse.

3.

Remove the blown fuses and replace them with two new fuses
(catalog #900-8935).

4.

Slide each fuse holder into the power entry module until it snaps in place.

Fuse holders