Recalibrating Your Imager

User Guide | 189

For any other changes to the optical pathway, perform a flat field calibration by

clicking Reset in the Instrument Calibration > Flat Field panel and following the

on screen instructions.

2.

The software prompts you to restart the calibrations needed for the new
illumination sources.

3.

Click OK to exit the dialog box. Your settings persist until you make further
changes.

Note: If the Gel Doc XR+ and ChemiDoc XRS+ systems are used to conduct

quantitation with Bio-Rad’s stain-free gel technology, Bio-Rad recommends

performing the flat field calibration with the orange fluorescence reference plate

(catalog # 170-8008) after installing Image Lab 5.1. For instructions on

performing this calibration, see

Calibrating the Imager to Use the Orange

Fluorescence Reference Plate on page 242

.