Step 1. gel imaging for multichannel protocols, Step 1. gel imaging for multichannel – Bio-Rad Image Lab™ Software User Manual

Page 74

Advertising
background image

4 | Acquiring Images

74 | ChemiDoc MP Imaging System with Image Lab Software

Manual Exposure — use this setting to manually override automated

imaging. Exposure time can range from 0.001 to 7,200 seconds with the

ChemiDoc MP imager.

After imaging a gel with automatic exposure optimization, the display

provides a reference point from which to set your manual exposure time.

Note: You can view the exposure time of the image in the Image Info

window (see

Image Info on page 109

).

Step 1. Gel Imaging for Multichannel Protocols

You must configure acquisition settings for at least two channels to create a valid

multichannel protocol. Each channel can be assigned a color (red, green, or blue) so

you can easily identify each channel. Configure each channel separately.