Rice Lake MA37 Sartorius Moisture Analyzer User Manual

Page 10

Advertising
background image

10

Startup

Operating Instructions Moisture Analyzer MA37

Opening and Closing the Sample Chamber

t

Open or close the sample chamber by hand . Always open the heating module

fully .

Power Connection (Power Supply)

– Using the wrong power cables may cause fatal electric shock and damage
the equipment .

– Never plug the power cable into the socket when it is disconnected from the
analyzer (danger of electrical shock) .

– Only connect the analyzer to an AC socket with a protective grounding
conductor .

– Only standard-compliant extension cords with protective grounding conductors
are permitted for use with the analyzer .

– Intentionally disconnecting the analyzer from the protective grounding
conductor is not permitted .

t

Check the plug design of the power cable .

If the plug design of the power cable does not comply with your country’s
standard, please inform the nearest Sartorius representative or your dealer .
The power must be connected in accordance with the regulations applicable
in your country . In order to connect the equipment to the power supply
(protection class 1), use a correctly installed wall outlet with a protective
grounding conductor (PE) and a fuse of a maximum 16 A . The power plug or
another suitable disconnecting device for the power must be easily accessible
so that the analyzer can be disconnected from the AC power supply in the
event of an emergency .

t

Connect the analyzer to the AC power .

t

Position the cable so that it cannot be damaged or hinder the measurement

process .

Safety Precautions

If you use an electrical outlet that does not have a protective grounding
conductor, ensure that an equivalent protective conductor is installed by a
certified electrician (as specified in the applicable regulations for installation in
your country) . The protective effect must not be negated by using an extension
cord without a protective grounding conductor .