MA Performance Test

Operating Instructions Moisture Analyzer MA37

+

4.000

g

Place pan, close hood for taring

D

t

Place a sample pan on the analyzer .

t

Close the hood .

y

The moisture analyzer tares automatically .

Open hood

0.000

g

D

t

Open the hood .

t

Open the packaging of the ReproEasy pad (accessory) and remove the pad .

t

Remove the protective wrapping .

t

Affix the pad to the center of the sample pan, sticky side down .

Fill in reference sample and

close hood to start

;

+

2.513

g

t

Place the ReproEasy pad reference sample (accessory) on the sample pan .

Close hood to start

;

+

6.721

g

t

Close the hood .

y

The performance test starts automatically .

During the performance test, the process status light on the analyzer flashes,
if the light has not been switched off in the menu .