Rice Lake MA37 Sartorius Moisture Analyzer User Manual
Page 40
40
MA Performance Test
Operating Instructions Moisture Analyzer MA37
+
4.000
g
Place pan, close hood for taring
D
t
Place a sample pan on the analyzer .
t
Close the hood .
y
The moisture analyzer tares automatically .
Open hood
0.000
g
D
t
Open the hood .
t
Open the packaging of the ReproEasy pad (accessory) and remove the pad .
t
Remove the protective wrapping .
t
Affix the pad to the center of the sample pan, sticky side down .
Fill in reference sample and
close hood to start
;
+
2.513
g
t
Place the ReproEasy pad reference sample (accessory) on the sample pan .
Close hood to start
;
+
6.721
g
t
Close the hood .
y
The performance test starts automatically .
During the performance test, the process status light on the analyzer flashes,
if the light has not been switched off in the menu .