Care and Maintenance

Operating Instructions Moisture Analyzer MA37

Contaminated Equipment:
– Health risk from product contamination due to product deposits and the

collection of residue with microbial contamination .

– Health risk from biological or microbiological substances .
– Observe cleaning specifications .
– Check the cleaning result closely .

Cleaning the Heating Module

Danger of Burns:
The inner part of the heating module and parts of the sample chamber may be
extremely hot .
Wait until the heating module has cooled completely .

Avoid touching the heating element . To remove spray residues or deposits from
the heating element, use a weak solvent such as ethanol .

Unlocking the Heating Module
1) Open the heating module all the way .
2) Pull the unlocking handle at the back of the heating element to unlock the

heating element .

Removing the Heating Module
3) Pull the heating module up out of the guide .