Moisture Analysis

Operating Instructions Moisture Analyzer MA37

Moisture Analysis

Purpose: The moisture analyzer can be used to determine the moisture content of liquid,

pasty, or solid samples .

Accessories: Depending on the consistency of the sample, one or two filters are recommended in

order to distribute the sample optimally in the sample pan .
– Liquid sample: Tare one filter with the sample pan, then distribute the sample

drop by drop onto the filter .

– Pasty sample: Tare two filters with the sample pan, then place the sample

between the two filters and press together evenly .

– Solid sample: Distribute the granulated sample evenly onto the sample pan

without a filter . Use one or two filters for solid samples that liquefy at high
temperatures or contain fat or sugar, in order to prevent crusting .

Danger of poisoning or caustic burns!
Substances that contain poisonous or caustic ingredients may produce poisonous
gases when dried, which can cause irritation (to the eyes, skin, or airways) or nausea
and lead to serious injury or death .
– These samples must only be dried in a laboratory hood .

Corrosion due to aggressive vapors!
Substances that emit aggressive vapors when heated may cause condensation to
form on cooler housing parts and lead to corrosion .
– Use small sample quantities of these substances .

Viewing and Changing Method Parameters

Moisture in %M

Start new measurement

Standard

Fully automatic

105

°C

Wednesday, 2014-01-08

16:50

Before starting the moisture analysis, you can view the current parameters and

change them if necessary .
– A method configuration for moisture analysis is preset . You can change and

save the parameters to suit your requirements .

The most important parameters (method name, temperature, switch-off criterion)
are displayed .

t

To check or view additional parameters, select the gray button at the bottom of

the main screen on the display .

Customize parameters

Print

No. of filters

Print lines

0

Off

Standby temperature

Off

y A summary of the current parameters is displayed .

If a lab printer is connected, you can print out the method parameters:

t

Select the print symbol .

y The parameters are printed out .

t

Select

o

to close the display .

t

To change the parameters, select the menu key at the bottom left of the main

screen on the display .