Calibration and Adjustment

Operating Instructions Moisture Analyzer MA37

Settings

Language
Date and time
Device information
Calibration/Adjustment
Printout

y The

Settings

menu appears .

t

Select

Calibration/Adjustment

.

Calibration/Adjustment

MA performance test
Weighing system adjustment
Calibration report

t

Select

Weighing system adjustment

.

t

Select

CAL-Extern

to calibrate the analyzer using an external calibration

weight .

y

The

Calibration/Adjustment

window appears .

t

Close the hood and select

l

to confirm .

t

The

CAL-Extern

window appears .

t

If necessary, please zero the integrated balance .

y

A message appears on the display, prompting you to place the calibration

weight (50 g) on the analyzer .

t

Place the calibration weight on the pan support .

y

Calibration starts automatically .

When calibration is complete, a report appears on the display . The report gives
the deviation detected during calibration . You are asked whether the weighing
system should be adjusted . After adjustment the report shows the result of the
adjustment .

t

Close the report: Select

<

.

y

The moisture analyzer is now calibrated . If you have also carried out an

adjustment after calibration, the moisture analyzer is also now adjusted .

CAL-Extern

START

Please set balance to zero
G - 0.401 g

CAL-Extern

Please put on required weight

Close hood

Stat - 50,000 g