System Settings (Setup Menu)

Operating Instructions Moisture Analyzer MA37

Device Information

The manufacturer, model, serial number and software version of the moisture
analyzer are displayed here .

Calibration/Adjustment

Calibration/Adjustment

MA performance test
Weighing system adjustment
Calibration report

The settings for calibration and adjustment of the moisture analyzer can be defined

here .

Under

MA performance test

:

Test functions for different temperature ranges are available here . You can only use
these functions with the ReproEasy pad (accessory) .

Under

Weighing system adjustment

:

t

Select the option of your choice .

CAL-Extern

: Carry out calibration with an external calibration weight

(see Chapter “Calibration and Adjustment”) .

Weighing only

: Use the weighing function to use the moisture analyzer

as a simple lab balance (see Chapter “Calibration and Adjustment”, section
“Weighing”) .

Under

Calibration report

:

Here you can access the most recent calibration reports for the MA performance
tests and weighing system adjustments . Up to 99 reports are saved per day .
Calibration reports older than 30 days are deleted .

Printout

You can configure the settings for printing and data output here . Some of
the settings depend on the configuration of the USB interface (see Chapter “USB
Interface” on page 30) .

Under

Printout measurement

:

t

Select the settings for printing out the results on the connected lab printer .

Printout measurement

: The printout can be activated and deactivated

during the measurement here .

Interim results

: If the printout is activated, you can also activate the

output of intermediate values during the measurement here .

Interval [sec]

: Enter the desired interval in seconds for the output of the

intermediate results .

Under

Printout and report format

:

t

Configure all settings for the output format .

Result only

: Only the result of the moisture analysis is output .

With GLP and method parameters

: The GLP data and parameters used for

the moisture analysis are also output .

Device information

Manufacturer:

Sartorius

Model:

MA37

Serial number:

0034002458

Version BAC:

00-53-01.01 CN:1701

Version APC:

01-80-01.05 CN:0000

Version DOC:

07-40-03.00 CN:0000