System Settings (Setup Menu)

Operating Instructions Moisture Analyzer MA37

Under

Decimal separator

(also available when “PC spreadsheet format” has been

selected):

Decimal point

(default setting): The number value is transmitted with a

decimal point to the PC program (example: 99 .963 g) .

Decimal comma

: The value is transmitted with a decimal comma to the PC

program (example: 99,963 g) .

Under

Keyboard emulation

(also available when “PC spreadsheet format” or

“PC text format” has been selected):

Universal (Num Lock on)

(default setting): Data is sent as special keyboard

characters in ASCII format (ALT + number pad) .
Requirement: Numbers lock must be turned on on the PC keyboard .

English (USA)

: Data is sent according to a keyboard set to English (USA) .

Requirement: For PC applications, such as MS Excel, the keyboard must be set
to English (USA) .

You can find more information in Chapter “USB Interface” .

2) Connection to Sartorius Lab Printer via the Supplied Printer Cable

Under

Device/Protocol

:

t

Select the desired setting for the connection to the printer .

YDP30/YDP40

: The connected printer is detected automatically and the

connection is established .

Off

: The connection to the printer is deactivated .

3) Connection to a Serial Printer or Another External Serial Device via a 9-pin

Serial Interface (USB Mini-A to RS-232)

Under

Device/Protocol

:

t

Select the desired setting for the connection to the device .

Printer YDP20-0CE

USB-RS232 SBI

Off

: The connection is deactivated .

Check the settings for

Printer YDP20-0CE

:

t

Set the printer to the default setting: 1200 baud, 7 bits, odd parity,

1 stop bit, CTS/RTS handshake

Under

RS-232 Configuration

:

This option can be selected if under

Device/Protocol

the setting

USB-RS232 SBI

has been selected .

t

Select the desired setting for the RS-232 interface .

Baudrate

: 600 to 19,200 (default setting: 9600)

Databits

: 7 bits or 8 bits (default setting: 8 bits)

Parity

: Odd, even, or none (default setting: odd)

Stopbits

: 1 bit or 2 bits (default setting: 1 bit)

Handshake

: Software (XON, XOFF), hardware (CTS, RTS), or off

[default setting: hardware (CTS, RTS)]