42

Calibration and Adjustment

Operating Instructions Moisture Analyzer MA37

Calibration and Adjustment

Background During

calibration, a calibration weight is used to determine how much the

displayed value deviates from the actual measurement value . This deviation is
compared against a preset target value . The subsequent

adjustment corrects

this deviation or reduces the permissible error limits .
Calibration and adjustment are combined in one process in the moisture analyzer .
After each calibration the analyzer is automatically adjusted .

When and How Often The moisture analyzer must be calibrated and adjusted, for example:

– When the ambient conditions have changed (temperature, humidity, or air

pressure)

– When the moisture analyzer is set up in a different location or moved in its

current location .

An external calibration weight is required for calibration and adjustment (100 g) .
Please note the tolerance of the calibration weight being used .

Calibrating the Moisture Analyzer

t

Make sure that the pan support is empty .

t

Select the menu key on the main screen .

Moisture in %M

Standard drying

105 °C

Standby temperature

-

Initial weight

-

End

Fully automatic

Target value

-

No automatic printout

y The menu is displayed .

t

Select the

i

(Setup) key in the menu .