Error Messages

Operating Instructions Moisture Analyzer MA37

Calibration/Adjustment Error Messages

“Weight is too low .”

When a weight that is too light is placed during
external calibration .

“Weight is too high .”

When a weight that is too heavy is placed during
external calibration .

“Not able to save cal . data .” When the memory is full .

Information about Analyzer Maintenance

“Maintenance interval
expired .”

When the maintenance date set by the Sartorius
Service Center has been exceeded .

USB Device Error Message

“The attached USB device is
not supported .”

When a USB device (printer) that is not permitted by
Sartorius has been connected .

Problems

Problem/Meaning

Remedy

Selected temperature is too high and the
sample is oxidized;
sample is boiling or burning and the
splashes are continuously changing the
weight .

– Reduce drying temperature
– Lay glass-fiber filter onto the sample
– Reduce the sample quantity or apply

the sample more evenly

– Select a semi-automatic switch-off

criterion or select timer mode

Measurement time too long

– Increase temperature
– Reduce sample quantity
– Preheat the analyzer by running a

drying process for 2 or 3 minutes
with an empty sample pan, for
example

Sample loses weight before it is analyzed

– Remove the sample pan and apply

the sample outside the analyzer

Sample is liquid or pasty

– Use glass fiber filter

Sample has only a low moisture content

– Increase sample quantity

Insufficient heating output

– Clean temperature sensor

Place of installation unstable
(vibrations, etc .)

– Set up the scale in another area