5

Safety Information

Operating Instructions Moisture Analyzer MA37

Note the IP protection class of the analyzer! Do not allow liquid penetration .
The protection class specifies the suitability of equipment for various environmental
conditions (moisture, foreign bodies) .

Before cleaning the analyzer: Unplug the power supply cable from the socket .

Position the power cable so that it cannot touch the hot surfaces of the analyzer
and is not a trip hazard . Disconnecting the grounding conductor is prohibited .

The device may only be opened by specialized personnel trained by Sartorius .

Danger due to Fire or Explosion!

– Flammable or explosive substances
– Substances containing solvents
– Substances that release flammable or explosive gases or vapors during the drying

process: When working with these types of samples, use a drying temperature
that is low enough to prevent a fire or explosion from occurring .

– If in doubt, carry out a careful risk analysis .
– Wear safety glasses .
– Always work with small sample quantities .

Never leave the analyzer unattended in these cases.

No modifications are permitted to the analyzer .

Symbols Used on the Device

When setting up the moisture analyzer, leave enough space to prevent heat from
building up and to keep the analyzer from overheating:
– 1 m above the analyzer
– 20 cm around the analyzer

The vent above the sample must not be covered or modified in any other way .
During operation do not open the hood, because the heating unit may be very hot .

Do not put any flammable substances on, under, or near the analyzer because
the area around the heating unit will heat up .

Be careful when removing the sample: The sample itself, the heating unit, and the
sample pans may still be extremely hot .

Observe the additional safety and danger information in the following chapters .