Calibration and Adjustment

Operating Instructions Moisture Analyzer MA37

Viewing the Calibration Report

Calibration/Adjustment

MA performance test
Weighing system adjustment
Calibration report

t

To view the most recent calibration reports: Go to the

Calibration/Adjustment

menu and select the

Calibration report

submenu .

Here you can select the most recent reports for the MA performance tests
or weighing system adjustments and then access the most recent calibration
reports, which are sorted by date .
Up to 99 entries per day are saved . Calibration reports older than 30 days are
deleted .

y

Menu Calibration report .

Example of a Calibration Report:

--------------------
2014-07-29     12:02
    Sartorius
Mod.          MA37-1
SerNo.    0031710609
BAC:        00-53-01
APC:        01-80-01
DOC:        07-40-03
--------------------
2014-07-29 

     12:02

External calibration
Start: manually
Set   +   50.000 g
Dev   -    0.002 g
External adjustment
Dev        0.000 g
Temp  +     23.7  

°C

--------------------
Name:

--------------------

Calibration Report

MA performance test
Weighing system adjustment