Moisture Analysis

Operating Instructions Moisture Analyzer MA37

Customize parameters

Initial weight

Off

Result display

Target value

Standby temperature

Moisture in %M

Off

Off

t

Scroll through the list if necessary and change further parameters .

t

When you are finished making changes, select

o

.

Save changes

Temperature: 110 °C
Initial weight: On
Target value: On

NO

YES

y The changed method parameters are displayed .

t

Select the desired option:

– Make further changes: Select

<

. This takes you back to the list of

parameters .

– Reject changes: Select

NO

.

– Save changes: Select

YES

.

y

The menu is displayed .

t

To return to the main screen, select

k

.

Performing the Moisture Analysis

Moisture in %M

Start new measurement

Standard

Fully automatic

105

°C

Wednesday, 2014-01-08

16:50

When all parameters have been set to suit your requirements, you can start the

moisture analysis .

Observe the following instructions in order to ensure successful measurement:
– Wear gloves for each measurement, so that the liquid in the sample is not

influenced by handling .

– Remove the sample directly from the process or keep it in an airtight and

watertight container until measurement .

– Ensure that the sample is representative for the product and is homogeneous

(shake, stir, or mix before measurement) .

– If using a filter or filters, store the filters in airtight and watertight packaging .

t

Select

START

on the main screen or open the hood .

If you selected

START

on the main screen:

Open hood

0.000

g

D

t

Open the hood now .

y

The next step is displayed when the hood is opened .