MA Performance Test

Operating Instructions Moisture Analyzer MA37

MA performance test

Device is tested with reference sample.

Temperature: 70 °C

Duration: 05:00 min

00:13 min

y The display shows the progress of the test and the remaining time .

Performance Test Measurement Not Successfully

Completed

t

MA performance test

Measurement canceled

y The message “

Measurement canceled

” appears .

t

Select

l

to confirm .

y

The display switches back to the menu .

If a fault has been detected:

t

MA performance test

Device is not OK.

5 MApi << 10 MApi

Temperature 70 °C

Please contact Sartorius service

a

y The message “

Device is not OK

” appears .

t

Select

m

to confirm .

Contact Sartorius service if the message “

Device is not OK

” appears .

Performance Test Successfully Completed

a

l

MA performance test

Device is OK.

Result 10 MApi

Temperature 70°C

3

y If the test is successful, the message “Device is OK” first appears . Then the

analyzer indicates that the sample can be removed .

Remove old sample

t

Select

l

to confirm .

y You can remove the sample and return to the menu .