background image

36

Moisture Analysis

Operating Instructions Moisture Analyzer MA37

+

4.000

g

Place pan and two filters, close

hood for taring

D

t

Place an empty sample pan on the analyzer, along with one or two filters if

necessary .
The number of filters depends on how many filters are set in the parameters for
the current method .

t

Close the hood .

y

The moisture analyzer tares automatically .

Open hood

0.000

g

D

t

Open the hood .

Fill in sample and close hood to

start

;

-

0.000

g

t

Place the sample on the sample pan or filter(s) .

Fill in sample

;

+

1.929

g

2.000

y If you have activated the “Initial weight” option in the method parameters,

a target value display appears .

t

Add the sample to the sample pan slowly . When the sample quantity reaches

the preset range, the target value display turns green .

Close hood to start

;

+

1.929

g

2.000

t

Close the hood .

y

The moisture analysis starts automatically .

During the measurement, the process status light on the analyzer flashes,
if the light has not been switched off in the menu .