Apply the sample to the sample pan:

t

t

Apply the sample to the sample pan in a thin, even layer (height: 2 to 5 mm,
weight: 5 to15 g); otherwise:

– A sample applied unevenly will result in a nonuniform distribution of heat

– A sample will not be dried completely

– The analysis time will be unnecessarily extended

– The sample will burn or a crust/skin will form on its surface as a result of a

very thick layer

– The crust makes it difficult or impossible for moisture to escape from the

sample during the drying process

– An uncertain and unknown quantity of moisture will remain in the sample

t

t

Apply liquid samples, pasty samples, transparent samples, or samples that can
melt to a glass-fiber filter (order no . 6906940); this provides the following
benefits:

– Uniform distribution due to the capillary effect

– Liquids are prevented from beading together and forming drops

– With larger surfaces, the moisture can evaporate faster

– Colorless/transparent samples do not reflect

– Considerably more convenient than the “sea-sand method”

When drying samples containing sugar, a crust or skin can form and seal the
surface . A glass-fiber filter is particularly useful in such cases . The moisture can
evaporate downwards through the surface of the filter . Crust/skin formation can be
prevented or limited by placing the glass-fiber filter on top of the sample .

t

t

Cover solid, heat-sensitive samples with a glass-fiber filter (order no . 6906940);
this provides the following benefits:

– Gentle heating, because the sample surface is shielded from excessive heat

– Higher temperature setting can be selected

– Uniformity of the sample surface

– Fast evaporation of the moisture

– Excellent repeatability for samples containing fat

Heating Programs
Two heating programs are available for performing moisture analysis of a
substance:

– Standard drying

– Gentle drying

Standard drying:
For standard drying, you need to enter the final temperature . The final temperature
is reached using an overshoot if necessary . The maximum set temperature is 160°C .

t [s]

T [°C]

Standardtrocknung

17

Operating Instructions Moisture Analyzer MA37

Basic Principles