Starting the Analysis

– After closing the hood:

If the initial weight condition is met, the analyzer prompts you to close the
hood . The start weight is accepted as soon as the sample chamber is closed and
the measurement value is stable .

End of Analysis

– Fully automatic:

The analysis ends as soon as the weight loss per 24 s is below an automatically
detected threshold .

– Semi-automatic, absolute:

The analysis ends as soon as the weight loss per the set interval is lower than a
user-defined threshold (set in milligrams) . The weight loss is entered by the user .

– Semi-automatic, percentage:

The analysis ends as soon as the weight loss in percent is lower than a user-
defined threshold (set in percent) . The percentage is entered by the user .

– Time

The analysis ends as soon as the specified time has elapsed .

– Manual:

Switch off using the END key . The maximum analysis duration is 24 hours .

Weight

Moisture (steam)

Autom. stopping

Time

19

Operating Instructions Moisture Analyzer MA37