Specifications – Rice Lake MA37 Sartorius Moisture Analyzer User Manual

Page 62

Advertising
background image

62

Specifications

Operating Instructions Moisture Analyzer MA37

Specifications

Dryer Functions

MA37-1

Temperature range and settings

40°C–160°C, in increments of 1°C; stand-by temperature selectable from 40–100°C

Sample heating

Infrared heating using a metal tube heater

Heating programs

Standard drying, gentle drying

Switch-off criterion

Optional:

– Fully automatic

– Semi-automatic mg (1–50 mg/5–300 s)

– Semi-automatic % (0 .1–5 .0%/5–300 s)

– Time (2:00–99 .59 min .)

– Manual

Displayed result

Moisture in %M and g; dry weight in %S and g; ATRO in %M/S

Net weight, approx .

6 .3 kg

Weighing Functions

Max . weighing capacity

70 g

Repeatability, average

For initial sample weight approx . > 1 g: ± 0 .2%

For initial sample weight approx . > 5 g: ± 0 .05%

Readability

1 mg; 0 .01%

Typical sample quantity

5 to 15 g

External calibration weight

50 g (E2)

Disposable sample pan dimensions

d

90 mm

Interface

Mini USB: Automatic detection of Sartorius printers YDP30 and YDP40,

direct data transfer to Microsoft

®

Windows programs without additional software,

programmable data output interval

Power Supply
Input supply voltage

100–120 V AC and 210–240 V AC (automatic voltage detection), ±10%, 50/60 Hz

Protection class and standards

Protection class I as per EN 61010-1/IEC 61010-1; up to 3000 m above sea level; IP20 as per

EN 60529/IEC 60529

Power supply

Using country-specific power cable from Sartorius only

Power consumption

Max . 640 W (typically)

Ambient Conditions

The specifications are valid for the following ambient conditions:
Environment

For indoor use only

Ambient temperature

+10°C to +30°C

Operating temperature

Operation guaranteed between +5°C and +40°C

Storage and shipping

–10°C to +60°C

Height

Up to 3000 m above sea level

Relative humidity

15% to 80% for temperatures up to 31°C non-condensing,

decreasing linearly up to 50% relative humidity at 40°C and 20% at 50°C

Safety of Electrical Equipment

As per EN 61010-1/IEC 61010-1

Safety requirements for electrical equipment for measurement, control, and laboratory use —

Part 1: General requirements

Electromagnetic Compatibility

As per EN 61326-1/IEC 61326-1

Electrical equipment for measurement, control, and laboratory use — EMC requirements —

Part 1: General requirements

Defined immunity to interference:

Suitable for use in industrial areas

Limitation of emissions:

Class B [suitable for use in residential areas and areas that are connected to a low voltage

network that (also) supplies residential buildings] . Therefore, the

analyzer can be used in both areas .