MA Performance Test

Operating Instructions Moisture Analyzer MA37

MA Performance Test

The performance test can be used to check whether the analyzer has a fault
or is ready for operation . During a performance test, both the heating unit and
the weighing system are tested at the same time . You will need a ReproEasy
pad (accessory) in order to carry out a performance test .

Starting the Performance Test

Calibration/Adjustment

MA performance test
Weighing system adjustment
Calibration report

t

Select “MA performance test .”

y

The display switches to a view with three heating temperatures .

Please ensure that the device has not been heated for at least one hour
before a performance test is carried out .

t

MA performance test

The device is too warm for a performance

test.

The device is still cooling down.

m

t

Select “

m

” to confirm the message and start the performance test again later .

MA performance test

Temperature 70 °C
Temperature 100 °C
Temperature 130 °C

t

Select one of the suggested temperatures given for the method you are using .


Open hood

0.000

g

D

t

Open the hood .

y

The next step is displayed when the hood is opened .