Care and Maintenance

Operating Instructions Moisture Analyzer MA37

Care and Maintenance

Service
To ensure its continued weighing accuracy, we recommend performing regular
servicing on your analyzer at least once a year . The Sartorius Service Center offers
different service contracts for this purpose, which can be individually adapted to
suit your needs (see also www .sartorius .com/service) . Each service produces a
calibration certificate . The analyzer and its connections must undergo a technical
safety inspection on a regular basis carried out by an electrician (e .g ., every 2 years) .

Repairs
Repair work must only be carried out by trained service technicians . The analyzer
must be unplugged during repair work . Unplug the power cable from the socket .
Failure to do so may affect the measurement accuracy of your analyzer and pose
a serious risk to the user . Contact the Sartorius Service Center or a Sartorius dealer
to arrange for proper repairs .

Cleaning the Analyzer

Cleaning the Control Panel
Switch the display to standby mode to avoid modifying the settings for operation
during cleaning .

t

Select the menu key to switch to the method menu .

t

If you then select

j

, the display is switched off .

t

To switch the display back on: Select

j

at the bottom left of the display .

The moisture analyzer starts in the application most recently used before the
analyzer was last switched off .

Cleaning the Housing

Disconnect the analyzer from the power before cleaning . Disconnect any data
cables from the moisture analyzer .

Never open the analyzer housing . The parts contained in the housing cannot be
cleaned, repaired, or replaced by the operator .

– Make sure that no liquid or dust gets into the analyzer .
– Remove the disposable sample pan and pan support from the analyzer .
– Never use cleaning agents that contain solvents or abrasive ingredients, which

can ultimately damage the equipment .

t

Clean the outside of the housing with a lint-free cloth and mild cleaning agent

(e .g ., isopropanol) .

t

Then dry the analyzer with a soft cloth .