Error messages – Rice Lake MA37 Sartorius Moisture Analyzer User Manual

Page 53

Advertising
background image

53

Error Messages

Operating Instructions Moisture Analyzer MA37

Error Messages

Certain events cause a message to appear on the display:
1 . Some error messages are displayed for approx . 3 seconds

(e .g .,

CAL-Extern

: calibration canceled) .

2 . Some error messages are displayed until they are acknowledged

by pressing a key (e .g ., “Incorrect access code” in “Service access”) .

3 . Information messages are displayed until they are acknowledged

by pressing a key .

Deactivated Keys

To prevent operating errors, only relevant functions/keys are displayed depending
on the situation . This prevents most operating errors from occurring . The following
buttons are only available in specific circumstances:

A

Only if there is a measured value in the zero setting range before
external calibration/adjustment .

D

Only if there is a measured value larger or smaller than “zero .”

Error Messages during Operation

“Value is too small .”

When an entered value is too low for the parameter .

“Value is too large .”

When an entered value is too high for the parameter .

Error 46

When the max . heating temperature has been briefly
exceeded .

Error 401 to 407

Faulty heating element . Contact the Sartorius Service
Center .