ISO/GLP-compliant Printout

Operating Instructions Moisture Analyzer MA37

ISO/GLP-compliant Printout

Features

You can have the device information, device ID, and the current date printed
before (GLP header) and after (GLP footer) the values from the weighing series .
The following data is printed:

GLP header:
– Date, time at the start of a weighing series
– Manufacturer
– Model
– Model serial number
– Software version number of the device (BAC)
– Software version number of the display (APC)
– Software version number of the heating module (DOC)
– Device ID with max . 14 characters (if activated under system settings)

GLP footer:
– Field for signature

Configuration

To print out the ISO/GLP report, define the following system settings
(see Chapter “System Settings”, section “Printout”):

t

Activate ISO/GLP-compliant printouts:
In the

Printout

menu under

Printout and Report Format

, activate the

With GLP and method parameters

option .

Operation

t

Print out the report after moisture analysis: Select