Introduction 7

1.5 Specifications

Table 1-1. MSI-7300 Specifications

Accuracy

Rated accuracy: ±0.1% of capacity. Typical accuracy ±(0.1%+1d) of
reading. ‘d’ equals one displayable increment.

Resolution

Standard displayed resolution: 2000-2500 ‘d’ . Hi-Res mode increases
resolution to 5000-6250 ‘d’. Internal A/D resolution: 24 bits.

Overload

Power

Display

Operating
Temp

Operating
Time

Load Cell &
Enclosure

Load Cell

F1 and F2

Calibration

Auto Zero
Maintenance

Auto-Off
Mode

Units

Filtering

Totalization

Peak

Setpoints

Service
Counter

WARNING

The MSI-7300 has a safe mechanical overload of 200% of capacity.
Overloads greater than 200% may result in physical damage to the
link. The ultimate overload is rated to 500%-700% of capacity (see
Table 1-2). At ultima
te overload, structural failure and dropped
loads may occur. Dropped loads may cause serious personal injury
or death.

Safe link mechanical overload, 200% of capacity. Ultimate link overload,
>500% of capacity. Typical ultimate overload is 700% or greater. Note that
the Dyna-Link is designed to have a greater safety factor than the connecting
shackles which have a typical ultimate safety factor of 600%
Battery operated, see Table 1-2 for battery size and quantity per unit.
Alkaline cells can be replaced with rechargeable NiMH cells.
Five digit, 1.22 in (31 mm) numeric digits.
Six digit, 1 in (26 mm) on 100000 lb units and higher

- 40°F to +122°F (-40°C to +50°C), Rated accuracy range -10°C to +40°C.

>150 hours typical (Alkaline C cells) / >300 hours typical (Alkaline D cells).

NEMA 4/IP65 anodized aluminum, 2024-T351 Aircraft Quality.

2000

Bridge

Programmable multifunction buttons for use as TEST, TOTAL, UNIT, PEAK,
TARE, NET/GROSS, VIEW TOTAL, PRINT AND High Res mode.

Fully digital calibration from the front panel or through a computer interface.

Standard, can be disabled internally.

Prolongs battery life by turning the power off after 15, 30, 45, or 60 minutes
(operator determined) of no Dyna-Link activity.

kg, lb, Tons (US short ton), Metric Tons, kilonewtons (other Units available
with custom calibrations).

Selectable - OFF, Low (LO), Medium (HI-1), High (HI-2).

Standard - Press button or automatic; TOTAL weight up to 99999 X1000
units

Uses unfiltered faster reading of A/D, (>220 readings per second).

Two internal standard setpoints and two ultrabright LEDs on indicator panel

Two independent 16 bit registers; register 1 updated each time the force
exceeds 25% of capacity; register 2 updated each time the force exceeds
overload; when register 1 exceeds 16383 or register 2 exceeds 1023,
display reads “LCnt” for load cell counter; test function shows the two
readings in order.