26 MSI-7300 Operator’s Manual

5.0 Calibration

The Dyna-Link 2 is calibrated using standard precision test weights. It is required that

the weight used is at least 10% of full capacity in order to achieve rated accuracy. For

example, use at least a 500kg test weight to calibrate a 5000kg capacity unit. The
Dyna-Link 2 supports load cell linearization with up to four span points that can be

calibrated in any order. Usually only one cal span point is necessary and is sufficient

to reach rated accuracy.
When adequate test weights are not available, the Dyna-Link 2 can be calibrated using

a constant calibration which is referred to as C-Cal. To use C-Cal, a factory generated

C-Cal number must be known. MSI supplies original and replacement load cells for

the Dyna-Link 2 with the C-Cal value stamped on the serial number label.
There are three aspects of calibration:

• Standard calibration - used for maintenance and routine calibration.
• Initial calibration - used to setup both the capacity and resolution (d) of the Dyna-

Link. It differs from standard calibration only in the initial steps. Initial

calibration is performed after a calibration reset which completely erases the

calibration and setup memory.

• C-Cal - If C-Cal values are known, the Dyna-Link can be calibrated without

weights.

5.1 Calibration Menu

The calibration menu contains three items: Cal, C-Cal, and auto zero maintenance.

The following procedures start with entering into the Cal menu, or for an initial

calibration, resetting the Dyna-Link 2 and then going to the Cal menu.

&$/

&&D/

6HWXS

Calibration Setup Menu

Calibrate with Weights

Calibrate with

Calibration number

Calibrate Menu

Figure 5-2 - Standard Calibration or

Figure 5-3 - Initial Calibration (After Reset)

Figure 5-4

Figure 5-5

Figure 5-1 Calibration Menu