Figure 5-2 - standar, Xq/g  /r, Figure 5-2 standard calibration procedure – Rice Lake MSI-7300 Dyna-Link 2 User Manual

Page 31

Advertising
background image

Calibration 27

Procedure for the Routine Calibration of the MSI-7300

XQ/G



/R$'

lb



Blinks
Capacity

1)

With the power on, initiate
calibration by holding down
the

F2

and

ZERO

keys until

the display reads CAL.

The CAL setup menu
appears.

2)

Press

F1

to start the

calibration procedure.

3)

The display reads UnLd
(unload) indicating you
should remove all weight
from the link.

You can choose to

leave bottom fittings on the link as
long as they are always part of the
load train.

4)

Press

F1

. The 7300 sets the

zero calibration point.

5)

If the zero is in range, the
scale will display PASS.
Then LoAd1 is displayed.

6)

Load the link with a precision
test weight.

For highest accuracy,

a test weight of 10% of capacity or
more is recommended.

7)

Press

F1

. The 7300 flashes

the capacity and calibration
units. If you are loading the
scale with the capacity
weight, skip to step 10.

8)

To enter a calibration
weight other than
capacity, press

F2

. The

displays far left digit will
flash zero indicating that
a number should be
entered.

&$/



blinking

3$66

No weight
on link

Flashes
Zero

Lift Test
Weight

example capacity

Error Correction: If you input a wrong
value, press

ZERO

to step back one digit

and reenter.

ENTER/SELECT

F1

ENTER/SELECT

F1

ENTER/SELECT

F1

SCROLL

F2

Press

Simultaneously

F2

0

ZERO

F1

Figure 5-2 Standard Calibration Procedure