46 MSI-7300 Operator’s Manual

7.0

Appendix

7.1 Troubleshooting

Problem

Possible Cause

Solution

Display is blank when
POWER key is pressed

Discharged battery

Replace cells, or if using NiMH,
recharge

Defective battery

Replace

Corroded battery or battery
contacts

Clean contacts

Defective switch or circuit
board

Requires authorized service

Display does not function
properly or front panel
keys do not function
normally or Dyna-Link will
not turn off

Improperly loaded software

Reinstall software

Faulty circuit board

Requires authorized service

Loose connectors

Requires authorized service

Dyna-Link does not
respond to tension
changes

Out of calibration

Calibrate

Faulty load cell

Replace

Load cell connector

Check connector and wires

Display over ranges below
100% of capacity

Tared tension is added to
load to determine overload
point

Return to gross tension mode

Zero requires adjustment

Rezero the Dyna-Link

Too much tension/load has
been zeroed

Rezero the Dyna-Link

Display drifts

AZM (Auto0) is turned off

Turn AZM on

Rapid temperature changes
such as moving the Dyna-
Link from indoors to
outdoors

Wait until the Dyna-Link
temperature has stabalized

Displayed tension shows
larger error

Dyna-Link not zeroed before
load is lifted

Zero the Dyna-Link with no load
attached

lb/kg units causing confusion Select proper units

Requires recalibration

Recalibrate

Display reading not stable

Excessive vibration in crane
system

Increase filtering or increase ‘d’
in Cal

Excessive side loading

Improve load train symmetry

Load cell faulty

Check LC connections

Display toggles between
“Error” and “Load”

Tension exceeds capacity

Immediately reduce tension

Faulty load cell or wiring

Check LC and LC wiring.

Table 7-1. Troubleshooting