4 c-cal calibration, C-cal calibration, Figure 5-4 – Rice Lake MSI-7300 Dyna-Link 2 User Manual

Page 36: Xq/g

Advertising
background image

32 MSI-7300 Operator’s Manual

5.4 C-Cal Calibration

When adequate test weights are not available, the MSI-7300 Dyna-Link 2 can be

calibrated using a cal number calibration which is referred to as C-Cal. To use C-Cal,

a factory generated C-Cal number must be known. MSI supplies replacement load

cells for the MSI-7300 Dyna-Link 2 with the C-Cal value stamped on the serial

number label. When a calibration is preformed with test weights, a new C-Cal is

generated.

Important

The C-Cal number must be known prior to starting this
procedure. For a MSI-7300 with its original load cell, MSI prints
this number on the Calibration Record, the serial number tag. C-
Calibration reduces slightly the absolute accuracy of the system
and is intended for non-critical use only. For highest accuracy,
calibrate the MSI-7300 with precision test weights.

XQ/G



1)

With the power on, initiate
calibration by holding down
the

F2

and

ZERO

keys until

the display reads CAL.

The CAL setup menu
appears.

2)

Press

F2

to scroll to the

C–Cal selection.

3)

Press

F1

to start the C-Cal

calibration procedure.

3)

The display reads UnLd
(unload) indicating you
should remove all weight
from the link.

You can choose to

leave bottom fittings on the Link as
long as they are always part of the load
train.

4)

Press

F1

. The 7300 sets the

zero calibration point.

5)

If the zero is in range, the
scale will display PASS.
Then C-CAL is displayed.

6)

Press

F1

. The last known

C–Cal value is displayed.

If the offered C-Cal value is correct,
push

F1

and jump to step 10.

In this example we’ll enter 15200 as
an C-Cal value. You must use your
known C-Cal value, not 15200.

&$/

3$66

No weight
on link

Flashes
Zero

ENTER/SELECT

F1

ENTER/SELECT

F1

SCROLL

F2

&&$/

&&D/

lb



example C-Cal Value

ENTER/SELECT

F1

Press

Simultaneously

F2

0

ZERO

F1

Figure 5-4 C-Cal Calibration Procedure