8 2-units/ 5-units, Units/ 5-units, Kg lb – Rice Lake MSI-7300 Dyna-Link 2 User Manual

Page 20

Advertising
background image

1) Program a function key to Peak

Hold (P-Hld)

In this example we’ll use

F1

for Peak Hold.

2) Prepare the test stand and test

sample.

3) Press

ZERO

to zero out any

residual strain on the link.

4) Press

F1 PkHold

. Confirm that

the “Pk” annunciator is on.

A small jump in the reading may occur

depending on the stability of your test
stand.

5) Apply the test force. The F1 LED

will blink three times for every
new peak it detects.

6) When the test force is removed,

the Peak value can be recorded.

7) To run a new test, press

F1 PkHold

to clear the peak

value. Confirm the “Pk”
annunciator is off. Then repeat
steps 3-6.

Peak Hold

kg

lb

M

kN

Ton



kg

lb

M

kN

Ton

Pk

kg

lb

M

kN

Ton



kg

lb

M

kN

Ton

Pk

Example Peak Captured Value

Peak Hold

kg

lb

M

kN

Ton



kg

lb

M

kN

Ton

kg

lb

M

kN

Ton



kg

lb

M

kN

Ton

Peak Cleared Value

F1

F1

0

ZERO

See Section 4.2 for
Function Key setup

16 MSI-7300 Operator’s Manual

Figure 3-1. Capture Peak Force

3.8

2-UNITS/ 5-UNITS

The

Fx-2.Unit

key will switch the force units between pounds force and kilograms

force. Selecting the

Fx-5.Unit

setting will scroll through all available units: lb, kg,

tons (U.S. short), metric tons, and kilonewtons.