6 Operation

104

K-375/376/377 Operation Manual, Version B

Determination Parameters
With the determination parameters the Determination mode (either “Standard” or “Direct distillation”)
can be determined.
In addition two sets of result units (each set consisting of one mass- and one volume-related unit)
can be selected for the standard determination mode.

Click on “Determination mode“ and select either
“Standard“ for a determination with a previous
digestion or “Direct distillation“ for a determination
without previous digestion.
For a direct distillation an additional factor and
regression factor can be specified for result calcula-
tion.

NOTE
The result for a direct distillation is calculated
following the linear equation

y= a·x+b where a is

given by the direct distillation factor and

b by the

regression factor.

The four result units (mass and volume) for the
determination mode “Standard” can be adapted
by clicking on it and selecting a corresponding
unit from the result units list. (See chapter “6.9.1
Settings“)