8 Troubleshooting

148

K-375/376/377 Operation Manual, Version B

14’100

Titrator timeout

Titration endpoint not reached. Check dosing unit,
if enough titration solution is available or update
frimware.

14’101

Titrator measured value is out of range

Verify the buffers used for calibration, check the
electrode.

14’501

Dosing unit not ready, please check

Verify if the dosing unit is connected to the distillation
unit.

14’502

Dosing unit locked

Check the dosing unit and switch distillation unit off
and on.

14’503

Dosing unit not ready, no exchange unit

Check if dosing unit is connected to the distillation
unit

14’504

Dosing unit not ready, no dosino

Check if dosing unit is connected to the distillation
unit

14’505

Dosing unit overload

Call service

14’506

Dosing unit not ready, cock blocked

Dismantle the dosing unit (see chapter “Trouble-
shooting the dosing unit”)

14’602

Titrator stopped, maximal volume reached

Make sure the electrode is in good working condi-
tions, the correct titration solution is used and that no
air bubbles are in titrator hoses.

14’603

Titrator stopped, stop endpoint reached

Check the titrator and if enough titration solution is
available

14’604

Titrator stopped, stopPot reached

Check the titrator and if enough titration solution is
available

14’605

Titrator stopped, stop time reached

Check the titrator and if enough titration solution is
available

15’001

No sampler connected

Switch on sampler and check the connection cable

15’002

Sampler: Target not achieved

Try again or call service

15’003

Sampler: Connection lost

Check the connection cable

15’101

Sampler: Shield open

Close shield

15’102

Sampler: Crash detected

Try again or call service

15’103

Sampler: Tube not found

Put sample tube in place or call service

15’104

Sampler: Tube not released

Remove sample tube or call service

15’105

Sampler: Error 5, reserve (gibt es nicht)

Call service

15’106

Sampler: Error 6, reserve (gibt es nicht)

Call service

15’107

Sampler: Shield not locked

Call service

15’108

Sampler: Error reference position

Call service

15’109

Sampler zeroadjustment not possible
because x or y deviation is larger then 3mm
or sampler was not in reference position
before adjustment start

Try again or call service

15’110

Sampler: Position error X-axis

Call service

15’111

Sampler: Position error Y-axis

Call service

15’112

Sampler: Position error Z-axis down

Call service

15’113

Sampler: Position error Z-axis up

Call service