6 storage tank connection, Storage tank connection, 5 .6 storage tank connection – BUCHI KjelSampler K-377 User Manual

Page 52: Notice

Advertising
background image

5 Putting into operation

52

K-375/376/377 Operation Manual, Version B

5 .6

Storage tank connection

To connect the storage tanks, proceed as follows:

• Cut the solaflex tube into pieces to the appropriate length.

• Insert a PTFE suction tube into the solaflex tube.

• Push a EPDM sealing ring over the solaflex tube.

• Now fasten the tubes to the tank with the red screw cover.

The storage tanks should not be positioned higher than the device itself and not lower than
1 meter below the instrument.

Fig. 5.7: Tank connection

All pumps are self-priming, no overpressure is necessary at the tanks.

NOTICE

Risk of device damage by calciferous water or wrong rong connected tanks.
• Use only distilled water for the H

2

O storage tank to keep the steam generator maintenance-

free.

• Make sure that the tanks are connected correctly. If the wrong tank (e. g. reagent tank

containing NaOH) is connected to the pump labelled as „H

2

O“, the steam generator will get

damaged.