7 Maintenance

129

K-375/376/377 Operation Manual, Version B

7 .4 .2

K-376/K-377 dip tube and sealing cap

The sealing cap is an expendable part which needs to be replaced periodically. We recommend to
replace this

every half year according to the below instruction. During use the dip tube will change

its color from white to grey. This is caused by the steam and should not have an impact on the
results. Nevertheless we recommend to replace it every half year together with the sealing cap.

!

CAUTION

Risk of burns by hot surface.
• Always let the device cool down after operation before touching the dip tube, the sealing or

the transfer and steam hoses.

• Take the sample rack away from the

tray

• Move the sampler arm to the service

position. (System Preparation 
Sampler functions  Move to service
position)

• The dip tube can be pulled out of the

sampler arm from below.

• To replace the sealing cap, first remove

the two hose clamps from the red
steam hose and the white transfer
hose using a flat nose plier.

• Remove both hoses from the two

fittings of the sealing cap.