6 Operation

110

K-375/376/377 Operation Manual, Version B

Result units

Create a new unit with

NEW and store it under a

new, unique name.

Select a base unit for the new unit calculation
Selectable are:

g N/L, % N, g Pr/L or % Pr

(N = nitrogen content, Pr = protein content)

Enter a correction factor and/or correction offset
if necessary and select the preferred number of
decimal places.

(The results are calculated following the linear equa-
tion

y= a·x+b where a is given by the correction

factor and

b by the correction offset.)

Dosage volume in status view

Set the increment in

[mL/click] for the direct

dosing of H

2

O, NaOH and H

3

BO

3

within the status

view. (Per click on the corresponding dose button,
the specified amount in mL will be dosed.)