3 transfer connection k-376 (k-377) – k-375, 1 connecting the k-376 to the k-375, Transfer connection k-376 (k-377) – k-375 – BUCHI KjelSampler K-377 User Manual

Page 46: ቤባ ቢ ብ

Advertising
background image

5 Putting into operation

46

K-375/376/377 Operation Manual, Version B

5 .3

Transfer connection K-376 (K-377) – K-375

The transfer connection between the K-375 and the K-376 or K-377 sampler consists of two hoses,
the white transfer hose and the red steam hose.
Both hoses have to be connected to the K-375 as well as to the sampler (K-376 or K-377) and
secured with hose clamps. The K-376 is delivered with both hoses premounted to the device.

!

WARNING

Serious chemical burns by corrosives. Risk of burns by hot steam.
• Never operate the K-375 together with a sampler while the sample transfer and steam hoses

are missing, defective, or incorrectly mounted.

• Make sure there is always enough room for a free movement of the sampler arm – if the

sampler arm collides with any object during movement, the transfer hose and/or the steam
hose may break!

5 .3 .1

Connecting the K-376 to the K-375

1. Fix the transfer hose support a with the

screw on the valve b on the rear side of
the K-375.

2. Guide both hoses c and d through the

transfer hose support.

Advertising
This manual is related to the following products: