8 Troubleshooting

151

K-375/376/377 Operation Manual, Version B

8 .4

Eliminating errors of the titrator

The buret is not properly filled

Possible reasons

Action/remedy

The reagent bottle is empty.

Replace or refill the reagent bottle.

The hose is not immersed deep enough
into the reagent bottle.

Immerse the hose deeper into the bottle, or fill up reagents.

The buret is not properly locked.

Lock the buret.

Air bubbles in the titration system

Possible reasons

Action/remedy

The hose connections are not tight.

Check whether the hose has been pulled out of the threaded connec-
tion and screw it on manually.
Replace the hoses including the threaded connections.
Refill the buret

Titration solution is not titrated/dosed

Possible reasons

Action/remedy

The buret is not properly filled.

Perform initial filling.

The hose or the titration tip are wrinkled
or blocked.

Check for a clear passage through hose and titration tip and replace the
corresponding parts, if necessary.

Undissolved parts in the titration solution.

Filter or replace titration solution.