BUCHI KjelSampler K-377 User Manual

Page 111

Advertising
background image

6 Operation

111

K-375/376/377 Operation Manual, Version B

Peripherals

Select

Yes for all present peripherals:

• Level sensor for receiver waste container
• Level sensor for sample tube waste container
• Sampler
• Balance
• Chiller
• Printer

NOTE
Unlike the level sensors for the storage tanks of
H

2

0, NaOH and H

3

BO

3

, the level sensors for the

waste containers have to be activated!
If a chiller is connected to the K-375 it is strictly
necessary to select “Yes” for “Chiller used” in
order to have the cooling water valve permanently
opened.

If a sampler is connected to the instrument, select
the type (1 tray for K-376, 2 trays for the K-377)
and enable or disable the auto cleaning functionality
of the sampler.

If a balance is connected to the instrument, specify
the communication settings for the balance.
• the Baud rate
• the number of Data bits (7 or 8)
• the Parity (none, even or odd)
• the number of Stop bits (1, 1.5 or 2)

NOTE
For details about the single parameters, please
refer to the operation manual of the balance.