7 Maintenance

135

K-375/376/377 Operation Manual, Version B

1

1

• Remove the two screws a on the rear

side of the instrument. The position of
the sodium hydroxide pump is marked
on the housing with the imprinting
“NaOH”.

3

3

2

1

• Open the service door.

• The sodium hydroxide pump is the

lowermost of the three pumps situated
on the left side.

• Disconnect the two plug connectors

a. (Note for reassembly: The upper
one connects the green cable, the
lower one the brown cable!)

• Disconnect the tube on the front side b.

• Unscrew the two screws c from the

front panel and take the front panel
away. The pump becomes loose now
and can be replaced.

• Reassemble in reversed order.