7 Maintenance

125

K-375/376/377 Operation Manual, Version B

7 .3

Monthly maintenance

7 .3 .1

Calibrating the pump

It is recommended to calibrate the pumps with the same volume as used for the methods.
To perform the calibration a graduated cylinder is needed.

Example H

2

O pump

Path: System Preparation  Pump Calibration

• Select H

2

O for “Pump”

• Enter the quantity to be dosed (e. g. 50 mL) for the

parameter “Dose volume”. (The last measured vol-
ume is always shown under “Calibration volume”)

• Press

Start to start dosing H

2

O.

• Transfer the dosed volume into a graduated cylin-

der, measure the volume and enter the measured
volume in the “Calibration volume [mL]” screen.