6 Operation

88

K-375/376/377 Operation Manual, Version B

For samples of type

Control Blank,

press “Name” and enter a name for the control
blank result.

Press “Method” and select the method to be used
for the determination of the control blank from the
list of the available methods.

Press “Group” and select a result group for the
storage of the result from the list of the available
result groups. (It is also possible to create a new
result group using the

New Group button.)

Using the

BALANCE button the weight of samples

can be taken over from a connected balance:

• Select all samples

• Press

BALANCE – all blanks and control blanks

are automatically deselected (no weight is
needed for blanks)

• Place the first sample on the balance and press

Enter on the balance – the first weight is taken
over from the balance and entered into the first
checked sample in the list.

• Proceed with the next sample

• When all sample weights have been taken over,

the balance mode is automatically left.

If certain samples of a rack that is already in prog-
ress have to be determined immediately this can be
done using the

Split Rack functionality:

Press

PAUSE to stop the determination of the

sequence.

Select the samples to be determined immediately
and press

SPLIT RACK.