7 Maintenance

133

K-375/376/377 Operation Manual, Version B

7 .5

Yearly maintenance

7 .5 .1

Replacement of wear parts

Replace the following components:

• Seals including the sealing cap of the sampler and the splash protector seal.

• NaOH pump and boric acid pump (the other pumps as required).

• Dip tube.

• pH electrode (if required depending on sample throughput and maintenance of the pH electrode).

• Wave spring in the sampler arm.

• Hoses inside the distillation unit, especially the ones that make contact with steam, NaOH and