8 Troubleshooting

147

K-375/376/377 Operation Manual, Version B

10‘302

Cooling water flow too low

Assure higher flow rate or check parameters in
Settings/Peripherals/Cooling water settings

10‘303

Low pressure during distilling

System pressure is below 150 mbar. Check for leaks
or call service.

10‘311

Pump H2O has no current

Faulty water pump. Exchange pump or call service.

10’312

Pump NaOH has no current

Faulty NaOH pump. Exchange pump or call service.

10’314

Pump H3BO3 has no current

Faulty boric acid pump. Exchange pump or call
service.

12’001

Valve steam (Y1) out of order

Faulty valve or wire harness. Call service.

12’002

Valve cooling water in (Y5) out of order

Faulty valve or wire harness. Call service.

12’003

Valve sampler steam (Y6) out of order

Faulty valve or wire harness. Call service.

12’004

Valve sampler transfer (Y7) out of order

Faulty valve or wire harness. Call service.

12’005

Valve 5 (not used) out of order

Faulty valve or wire harness. Call service.

12’006

Valve waste sample tube (Y2) out of order

Faulty valve or wire harness. Call service.

12’007

Valve aspiration in (Y3) out of order

Faulty valve or wire harness. Call service.

12’008

Valve receiver (Y4) out of order

Faulty valve or wire harness. Call service.

12’009

Valve H2O injection (Y8) out of order

Faulty valve or wire harness. Call service.

12’010

Valve H2O sample tube (Y9) out of order

Faulty valve or wire harness. Call service.

12’011

Valve waste receiver (Y10) out of order

Faulty valve or wire harness. Call service.

13’001

27V power supply overcurrent

Faulty electronic board. Call service.

13’002

Ventilator power overcurrent

Fan short circuit. Call service.

13’003

Ventilator electronics blocked

Check for blockage or call service

13’004

Ventilator interior blocked

Check for blockage or call service

14’001

Titrator not ready

Check if all cables are plugged to the titrator, restart
the system or call service.

14’002

Titrator information (version)

Call service

14‘003

Titrator not started

Titrator error. Check function in System Preparation/
Buret function. Call service.

14‘004

Titrator not started, pH value too low

pH value is below the set endpoint. Verify the elec-
trode, dosing unit and boric acid.

14‘005

Titrator not started, pH value too high

pH value is above the set endpoint. Verify the elec-
trode, dosing unit and receiving solution.

14’006

Wrong titration direction

Ensure pH electrode is imersed in receiving solution
and verify if the right titration solution is used.

14’007

Titration speed over specification

Overtitration. Use lower concentrated titration solution
or reduce titration speed.

14’008

Titration speed over specification

Overtitration during back titration. Use lower concen-
trated titration solution or reduce titration speed.

14’010

Titrator module could not create service 11

Distillation unit needs to be switched off and on

14’011

Titrator module could not create service 21

Distillation unit needs to be switched off and on

14’012

Titrator module could not create service 41

Distillation unit needs to be switched off and on

14’013

Titrator module could not create service 3

Distillation unit needs to be switched off and on