9 preparing the system, 1 preparing the software, Preparing the system – BUCHI KjelSampler K-377 User Manual

Page 57: 5 .9 preparing the system

Advertising
background image

5 Putting into operation

57

K-375/376/377 Operation Manual, Version B

5 .9

Preparing the system

5 .9 .1

Preparing the software

In general it is recommended to check and adapt

all device settings, located under

HOME  Settings
previous to the first use of the instrument.

Following a selection of the most common settings to be adapted is provided:

Define regional settings
HOME  Settings  Regional settings
Choose device language, keyboard layout, and date and time format

Set date and time
HOME  Settings  Date and time
Set date, time, and time zone

Define user (optional)
HOME  Settings  User administration
Different users with specific user rights can be defined. As long as no user is defined, no user admin-
istration will be used. For more details refer to the section “6.3 User concept”.

Check peripherals
HOME  Settings  Peripherals
Make sure all connected peripherals are selected and configured.

Specify an import and export path for results and other data
HOME  Settings  Import and Export
Data can be exported either to a USB device or to a network data share. If a network data share
shall be used, a path needs to be specified.

Depending on your preferred applications and methods the following items have to be defined:

Volumetric solutions
HOME  Volumetric Solutions
Define all solutions that may be used for your applications.

Reference substances (optional)
HOME  Reference Substances
Specify the reference substances together with their theoretical values.

Method (optional)
HOME  Methods
Define a new determination method or modify an existing method if necessary.

Blank Corrections
HOME  Blank Correction
Determine the general behavior of the system with respect to the blank correction.