7 Maintenance

126

K-375/376/377 Operation Manual, Version B

The value shown for “Calibration volume” is updated.

NOTE
Repeat this procedure until the entered and the dosed volume correspond. An acceptable difference
at 50 mL is ±5 mL.
H

2

O and NaOH can be dosed into the sample tube and then poured into a graduated cylinder for

measurement.
The H

3

BO

3

can be dosed directly into the receiving vessel and then poured into a graduated cylinder.

7 .3 .2

Checking the distillate amount

• Carry out preheating (three times) so that the system is warm before performing this test.
• Create a new method using the following parameters:

H

2

O volume:

0 mL

NaOH volume:

0 mL

Reaction time:

0 s

Distillation mode:

Fixed time

Distillation time:

300 s

Stirrer speed distillation:

5

Steam output:

100 %

Titration type:

None

Aspiration sample tube:

Yes

Aspiration receiving vessel:

No

• Run the method with an empty sample tube and empty receiving vessel.
• Measure the distilled quantity in the receiving vessel using a graduated cylinder.

NOTE
The distillate amount with above parameters must be

130 mL.