7 Maintenance

141

K-375/376/377 Operation Manual, Version B

1

To adjust the distance between the sample
tube holder and the splash protector seals
① for an optimum contact pressure,

• slightly loosen the two screws

• insert the gauge

① into the gap

between the holder and the housing
(insert the gauge according to the
condition of the seal – for a new seal
insert the thick end of the gauge, for
a worn sealing use the thin end, or
the middle part)

• move the sample tube holder

upwards against the gauge

① and

tighten the two screws on the holder

• remove the gauge

• insert a sample tube and test the

connection of the sample tube for
tightness

• if necessary repeat the adjustment

using another step of the gauge

Fig. 7.4 Adjustment of the sample tube holder

NOTE
If the adjustment of the sample tube holder does not remedy the problem, it may be necessary to
exchange the sample tube holder and/or the splash protector seal.