7 Maintenance

120

K-375/376/377 Operation Manual, Version B

7 .1

Daily maintenance

7 .1 .1

Before sample determination

Remove the electrode from the storage cap (the cap can fixed to the electrode holder on the front

side of the housing) and place it into the receiving vessel.

• Rinse the titration hoses (Path: System Preparation  Buret functions  Buret function “Dose”).

• Calibrate the pH electrode with fresh buffer solution, see also chapter “6.6.1 System Preparation”.

Prime the system (Path: System Preparation  Priming), see also chapter “6.6.1 System Prepara-

tion”.

• Submerge the pH electrode into the measuring medium at least up to the diaphragm.

NOTE
All glass parts must be warm before the analysis begins. Therefore a preheating of the system is
necessary if more than 15 minutes pass between two determinations (Path: Operator, System prepa-
ration, Preheating).
Cleaning between samples is not necessary, unless sample deposits are found in the splash
protector. (Path: System Preparation
Cleaning).